FDA Inspection 978525 — BioSphere Medical S.A.

BioSphere Medical S.A. — FEI 3002808761

The FDA completed an inspection of BioSphere Medical S.A. on Jun 16, 2016 in Roissy En France. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 8 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Jun 16, 2016
Fiscal Year
2016
Product Type
Devices
Project Area
Compliance: Devices
Location
Roissy En France (France)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
328621 CFR 820.90(b)(1)Procedures for product review,disposition lack of/inadequate
367621 CFR 820.30(f)Design verification - documentation
369621 CFR 820.100(b)Documentation
405921 CFR 820.22Quality Audits - defined intervals
407021 CFR 820.30(g)Design validation - documentation
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.