Z-0002-2016 Class II Terminated

Recalled by Biosense Webster, Inc. — Irwindale, CA

FDA device recall Z-0002-2016 was initiated by Biosense Webster, Inc. on Sep 10, 2015 and is designated Class II. Reason for recall: Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of… The recall status is terminated (terminated May 13, 2016). Affected quantity: 1035 (U.S.).

Recall details

Product Type
Devices
Report Date
Oct 7, 2015
Initiation Date
Sep 10, 2015
Termination Date
May 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1035 (U.S.)

Product description

Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.

Reason for recall

Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.

Distribution pattern

Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.

Code information

Catalog #: C3SOUND, 10439236, 10439072, 10439011, 10438577

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0002-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.