Z-0016-2022 Class II Terminated

Recalled by Paltop Advanced Dental Solutions Ltd — Caesarea

FDA device recall Z-0016-2022 was initiated by Paltop Advanced Dental Solutions Ltd on Jul 26, 2021 and is designated Class II. Reason for recall: Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek… The recall status is terminated (terminated Sep 21, 2023). Affected quantity: 16.

Recall details

Product Type
Devices
Report Date
Oct 13, 2021
Initiation Date
Jul 26, 2021
Termination Date
Sep 21, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16

Product description

Scan Abutment SU, Cat. No. 30-70104 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide a digital registration of an implant or abutment platform. The Paltop Advanced Dental Solutions Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient s chewing function.

Reason for recall

Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.

Distribution pattern

US Nationwide distribution. No foreign distribution from US.

Code information

All lots manufactured prior to July 21, 2021.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0016-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Paltop Advanced Dental Solutions Ltd? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Paltop Advanced Dental Solutions Ltd: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Paltop Advanced Dental Solutions Ltd →

Source: openFDA record for Z-0016-2022. Report a data issue: hello@regdatalab.com.