Z-0018-2022 Class II Terminated

Recalled by Paltop Advanced Dental Solutions Ltd — Caesarea

Recall Details

Product Type
Devices
Report Date
October 13, 2021
Initiation Date
July 26, 2021
Termination Date
September 21, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
236

Product Description

Scan Abutment WP, Cat. No. 30-70106 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide a digital registration of an implant or abutment platform. The Paltop Advanced Dental Solutions Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient s chewing function.

Reason for Recall

Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.

Distribution Pattern

US Nationwide distribution. No foreign distribution from US.

Code Information

All lots manufactured prior to July 21, 2021.