Z-0025-2018 Class II Terminated

Recalled by Medtronic Navigation, Inc. — Louisville, CO

FDA device recall Z-0025-2018 was initiated by Medtronic Navigation, Inc. on Jul 27, 2017 and is designated Class II. Reason for recall: The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process. The recall status is terminated (terminated Aug 5, 2019). Affected quantity: 423 devices.

Recall details

Product Type
Devices
Report Date
Oct 25, 2017
Initiation Date
Jul 27, 2017
Termination Date
Aug 5, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
423 devices

Product description

VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip

Reason for recall

The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.

Distribution pattern

US nationwide distribution.

Code information

Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip Lots - 000212782059, 000212823919, 000212894256, 000212978141, 000212783784, 000212854585, 000212926008, 000212996201, 000212823883, 000212882619, 000212956163, 000213145362; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip Lots - 000212354781, 000212656015, 000212760821, 000212913269, 000212384009, 000212660280, 000212827856, 000212922387, 000212465594, 000212664497, 000212838837, 000212935278, 000212507737, 000212698642, 000212846727, 000213000097, 000212510629, 000212715926, 000212856596, 000213001255, 000212624409, 000212745633, 000212873863, 000213009483, 000212626439, 000212753561, 000212873873, 000213041517, 000212632885, 000212756095, 000212879480, 000213056030; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip Lots - 000212141557, 000212503634, 000212647830, 000212796559, 000212148368, 000212504463, 000212701679, 000212846726, 000212434500, 000212549456, 000212707035, 000212901580, 000212438405, 000212553428, 000212745660, 000212912501, 000212473141, 000212632884, 000212795235, 000212963182

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0025-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-0025-2018. Report a data issue: hello@regdatalab.com.