Z-0038-2017 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

FDA device recall Z-0038-2017 was initiated by RAYSEARCH LABORATORIES AB on Sep 1, 2016 and is designated Class II. Reason for recall: RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose… The recall status is terminated (terminated Jun 11, 2019). Affected quantity: USA: 522 units, OUS 846 units.

Recall details

Product Type
Devices
Report Date
Oct 12, 2016
Initiation Date
Sep 1, 2016
Termination Date
Jun 11, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
USA: 522 units, OUS 846 units

Product description

RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Reason for recall

RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline (External ROI) in RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0).

Distribution pattern

Nationwide distribution

Code information

Build Numbers: 2.5.0.144, 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.42, 4.3.0.14, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0038-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for RAYSEARCH LABORATORIES AB? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for RAYSEARCH LABORATORIES AB: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about RAYSEARCH LABORATORIES AB →

Source: openFDA record for Z-0038-2017. Report a data issue: hello@regdatalab.com.