Z-0053-2014 Class III Terminated

Recalled by Alere San Diego, Inc. — San Diego, CA

FDA device recall Z-0053-2014 was initiated by Alere San Diego, Inc. on Sep 23, 2013 and is designated Class III. Reason for recall: Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail… The recall status is terminated (terminated Sep 26, 2014). Affected quantity: 38.

Recall details

Product Type
Devices
Report Date
Oct 30, 2013
Initiation Date
Sep 23, 2013
Termination Date
Sep 26, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38

Product description

Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage¿ TOX Drug Screen Control 1 or Alere Triage¿ TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen Controls are to be used with the Alere Triage¿ TOX Drug Screen tests and Alere Triage¿ Meters to assist the laboratory in monitoring test performance.

Reason for recall

Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.

Distribution pattern

Worldwide Distribution - USA (nationwide) and countries of: South Africa and Canada.

Code information

Part Number 94413. Lot Number C2681A.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0053-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0053-2014. Report a data issue: hello@regdatalab.com.