Z-0062-2020 Class II Terminated
Recalled by GE Healthcare, LLC
— Waukesha, WI
FDA device recall Z-0062-2020 was initiated by GE Healthcare, LLC on Mar 19, 2019 and is designated Class II. Reason for recall: The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since… The recall status is terminated (terminated Mar 30, 2020). Affected quantity: 22 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 16, 2019
- Initiation Date
- Mar 19, 2019
- Termination Date
- Mar 30, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22 units
Product description
AIR Technology Anterior Array, Catalogue Numbers: (a) 5746694, (b) 5746695, (c) 5746696, a receive-only RF coil designed for use with GE 3.0T MRI systems.
Reason for recall
The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.
Distribution pattern
Worldwide Distribution - US Nationwide in the states of AZ, CA, MS, NY, WI, DE, and countries of France, Germany, Japan, Netherlands, Spain, Sweden,
Code information
Catalogue # UDI # Mfg. Lot or Serial # 5746694 0100840682138611111812132100339 339 5746694 0100840682138611111809122100324 324 5746694 0100840682138611111805312100311 311 5746694 0100840682138611111901022100340 340 5746694 0100840682138611111811302100334 334 5746694 0100840682138611111808302100323 323 5746694 0100840682138611111809132100325 325 5746694 0100840682138611111809192100326 326 5746694 0100840682138611111801292100303 303 5746694 0100840682138611111801232100301 301 5746694 0100840682138611111808152100319 319 5746694 0100840682138611111808272100320 320 5746694 0100840682138611111812042100335 335 5746694 0100840682138611111806072100313 313 5746694 0100840682138611111807182100315 315 5746694 0100840682138611111802122100307 307 5746694 0100840682138611111811092100328 328 5746694 0100840682138611111812072100337 337 5746694 0100840682138611111807232100316 316 5746695 0100840682138611111808282100321 321 5746695 0100840682138611111808292100322 322 5746696 0100840682138611111812132100338 338
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0062-2020
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about GE Healthcare, LLC →
Use it in ChatGPT
Get a free API key
How the connector works