Z-0065-2014 Class II Terminated
FDA device recall Z-0065-2014 was initiated by Varian Medical Systems, Inc. on Aug 9, 2013 and is designated Class II. Reason for recall: The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization. The recall status is terminated (terminated Apr 10, 2014). Affected quantity: 3 sets.
Recall details
- Product Type
- Devices
- Report Date
- Oct 30, 2013
- Initiation Date
- Aug 9, 2013
- Termination Date
- Apr 10, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 sets
Product description
Reason for recall
Distribution pattern
Worldwide distribution: US (nationwide) distribution in CA and in the country of India.