Z-0065-2014 Class II Terminated

Recalled by Varian Medical Systems, Inc. — Charlottesville, VA

Recall Details

Product Type
Devices
Report Date
October 30, 2013
Initiation Date
August 9, 2013
Termination Date
April 10, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 sets

Product Description

VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA

Reason for Recall

The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.

Distribution Pattern

Worldwide distribution: US (nationwide) distribution in CA and in the country of India.

Code Information

Lots #6078184-1, #6078184-2, #6078184-4