Z-0069-2020 Class II Terminated
Recalled by Becton, Dickinson And Company, BD Biosciences
— San Jose, CA
FDA device recall Z-0069-2020 was initiated by Becton, Dickinson And Company, BD Biosciences on Aug 19, 2019 and is designated Class II. Reason for recall: Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio… The recall status is terminated (terminated Apr 12, 2024). Affected quantity: 78.
Recall details
- Product Type
- Devices
- Report Date
- Oct 16, 2019
- Initiation Date
- Aug 19, 2019
- Termination Date
- Apr 12, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 78
Product description
BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes
Reason for recall
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Distribution pattern
U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
Code information
BD FACSLyric Flow Cytometer serial numbers: Z662878000003, Z662878000041, Z662878000042, Z662878000043, Z662878000044, Z662878000045, Z662878000046, Z662878000047, Z662878000049, Z662878000050, Z662878000051, Z662878000052, Z662878000053, Z662878000054, Z662878000055, Z662878000056, Z662878000057, Z662878000058, Z662878000059, Z662878000060, Z662878000061, Z662878000062, Z662878000063, Z662878000064, Z662878000065, Z662878000066, Z662878000067, Z662878000068, Z662878000069, Z662878000070, Z662878000071, Z662878000072, Z662878000073, Z662878000074, Z662878000075, Z662878000076, Z662878000077, Z662878000078, Z662878000079, Z662878000080, Z662878000081, Z662878000082, Z662878000083, Z662878000084, Z662878000085, Z662878000086, Z662878000087, Z662878000088, Z662878000089, Z662878000090, Z662878000091, Z662878000092, Z662878000093, Z662878000094, Z662878000095, Z662878000096, Z662878000097, Z662878000098, Z662878000105, Z662878000106, Z662878000107, Z662878000108, Z662878000109, Z662878000110, Z662878000111, Z662878000112, Z662878000113, Z662878000114, Z662878000115, Z662878000116, Z662878000117, Z662878000118, Z662878000119, Z662878000120, Z662878000121, Z662878000122, Z662878000123, Z662878000048.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0069-2020
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.