Z-0076-2017 Class II Terminated
FDA device recall Z-0076-2017 was initiated by Elekta, Inc. on Sep 29, 2016 and is designated Class II. Reason for recall: When creating 3D plans using either MU or Dose weighting modes, if the user changes the Physician's Intent Rx Dose and/or the number of fractions, and then modifies the wedge angle, the MU value is… The recall status is terminated (terminated Jul 16, 2021). Affected quantity: 365.
Recall details
- Product Type
- Devices
- Report Date
- Oct 19, 2016
- Initiation Date
- Sep 29, 2016
- Termination Date
- Jul 16, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 365
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - US to CA, FL, KY, MD, MO, MT, NJ, NC, PA, WA, WI. Internationally to Algeria. Antiqua & Barbuda, Argentina, Australia, Austria, Belarus, Botswana, Brazil, Bulgaria, China, Columbia, Croatia, Czech Republic, Ecuador Estonia, France, Germany, India, Iran, Israel, Italy, Japan, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Portugal, Romania, Russia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, and Zimbabwe.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0076-2017
- openFDA file date
- Nov 10, 2026