Z-0082-2014 Class II Terminated
FDA device recall Z-0082-2014 was initiated by Maquet Medical Systems USA on Sep 12, 2013 and is designated Class II. Reason for recall: It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will… The recall status is terminated (terminated Jun 7, 2017). Affected quantity: 63.
Recall details
- Product Type
- Devices
- Report Date
- Nov 6, 2013
- Initiation Date
- Sep 12, 2013
- Termination Date
- Jun 7, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 63
Product description
MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.
Reason for recall
It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
Distribution pattern
Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO
Code information
Part No. 70104.8012 Serial Nos. 90410440; 90410452; 90410455; 90410456; 90410457; 90410459; 90410469; 90410474; 90410480; 904104781; 90410482; 90410484; 90410498; 90410499; 90410500; 90410501; 90410502; 90410503; 90410521; 90410522; 90410524; 90410526; 90410527; 90410531; 90410533; 90410534; 90410535; 90410550; 90410551; 90410552; 90410568; 90410571; 90410572; 90410573; 90410575; 90410576; 90410577; 90410580; 90410581; 90410582; 90410584; 90410585; 90410592; 90410596; 90410599; 90410600; 90410605; 90410606; 90410607; 90410608; 90410609; 90410610; 90410611; 90410612; 90410626; 90410638; 90410640; 90410642; 90410644; 90410646; 90410647; 90410650; 90410651; 90410652; 90410653; 90410656; 90410659; 90410660; 90410661; 90410662; 90410663; 90410664; 90410665; 90410666; 90410667; 90410668; 90410670; 90410700; 90410708; 90410846; 90410750; 90410786