Z-0085-2025 Class II Ongoing

Recalled by LEICA BIOSYSTEMS NUSSLOCH GMBH — Nussloch

FDA device recall Z-0085-2025 was initiated by LEICA BIOSYSTEMS NUSSLOCH GMBH on September 4, 2024 and is designated Class II. Reason for recall: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and the… The recall status is ongoing. Affected quantity: 1 unit.

Recall Details

Product Type
Devices
Report Date
October 23, 2024
Initiation Date
September 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples

Reason for Recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI/DI 04049188066237, Serial Number 990