Z-0086-2025 Class II Ongoing

Recalled by LEICA BIOSYSTEMS NUSSLOCH GMBH — Nussloch

FDA device recall Z-0086-2025 was initiated by LEICA BIOSYSTEMS NUSSLOCH GMBH on Sep 4, 2024 and is designated Class II. Reason for recall: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and… The recall status is ongoing. Affected quantity: 13 units.

Recall details

Product Type
Devices
Report Date
Oct 23, 2024
Initiation Date
Sep 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product description

Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples

Reason for recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Distribution pattern

US Nationwide distribution.

Code information

a) Material Number 1491510C002, UDI/DI (01)04049188002945, Serial Numbers: 5359022006, 5367012006, 5913, 5996, 5065, 5625022007, 5467072006, 5901, 5108122004, 5330, 5269, 5667; b) Material Number 1491510C003, UDI/DI (01)04049188002891. Serial Numbers: 5537

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0086-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0086-2025. Report a data issue: hello@regdatalab.com.