Z-0098-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 21, 2020
- Initiation Date
- September 1, 2020
- Termination Date
- January 20, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product Description
AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0
Reason for Recall
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
Distribution Pattern
US distribution to DE, NC, OK, PA, and TX.
Code Information
9BDS1515 08D51544 OBDS1596 OBPS2864 9BDS1529 OBDS1545 OBPS2799 OBPS2884 9BPS2735 OBDS1551 OBPS2800 OBPS2903 9BPS2767 08D51590 OBPS2810