FDA device recall Z-0112-2016 was initiated by Elekta, Inc. on Sep 15, 2015 and is designated Class II. Reason for recall: Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities. The recall status is terminated (terminated Apr 4, 2017). Affected quantity: 154 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 14, 2015
- Initiation Date
- Sep 15, 2015
- Termination Date
- Apr 4, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 154 units
Product description
Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Reason for recall
Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.
Distribution pattern
Worldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.
Code information
Software Version 5.10.00