FDA device recall Z-0118-2013 was initiated by Elekta, Inc. on Aug 21, 2012 and is designated Class II. Reason for recall: During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and… The recall status is terminated (terminated Nov 3, 2014). Affected quantity: 52 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 31, 2012
- Initiation Date
- Aug 21, 2012
- Termination Date
- Nov 3, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 52 units
Product description
Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.
Reason for recall
During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
Distribution pattern
Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
Code information
None as the product is software