Z-0118-2013 Class II Terminated

Recalled by Elekta, Inc. — Norcross, GA

FDA device recall Z-0118-2013 was initiated by Elekta, Inc. on Aug 21, 2012 and is designated Class II. Reason for recall: During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and… The recall status is terminated (terminated Nov 3, 2014). Affected quantity: 52 units.

Recall details

Product Type
Devices
Report Date
Oct 31, 2012
Initiation Date
Aug 21, 2012
Termination Date
Nov 3, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52 units

Product description

Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.

Reason for recall

During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.

Distribution pattern

Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.

Code information

None as the product is software

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0118-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0118-2013. Report a data issue: hello@regdatalab.com.