Z-0122-2016 Class III Terminated

Recalled by Alere San Diego, Inc. — San Diego, CA

FDA device recall Z-0122-2016 was initiated by Alere San Diego, Inc. on Aug 27, 2015 and is designated Class III. Reason for recall: Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect. The recall status is terminated (terminated Dec 22, 2015). Affected quantity: 190 kits.

Recall details

Product Type
Devices
Report Date
Oct 21, 2015
Initiation Date
Aug 27, 2015
Termination Date
Dec 22, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
190 kits

Product description

Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG) and glucose (GLU) test procedures.

Reason for recall

Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.

Distribution pattern

Nationwide Distribution.

Code information

Lot C3091 and Lot C3091A

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0122-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0122-2016. Report a data issue: hello@regdatalab.com.