Z-0126-2027 Class I Ongoing

Recalled by Ventec Life Systems, Inc. — Bothell, WA

FDA device recall Z-0126-2027 was initiated by Ventec Life Systems, Inc. on Sep 17, 2026 and is designated Class I. Reason for recall: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. The recall status is ongoing. Affected quantity: 1,532 units.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Sep 17, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,532 units

Product description

VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

Reason for recall

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Distribution pattern

Distribution US nationwide including Puerto Rico.

Code information

All Serial Numbers: UDI: 00855573007747