Z-0133-2027 Class I Ongoing
FDA device recall Z-0133-2027 was initiated by Ventec Life Systems, Inc. on Sep 17, 2026 and is designated Class I. Reason for recall: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. The recall status is ongoing. Affected quantity: 25,112 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2026
- Initiation Date
- Sep 17, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25,112 units
Product description
V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Distribution pattern
Distribution US nationwide including Puerto Rico.
Code information
All Serial Numbers: UDI: 00855573007822