Z-0132-2026 Class II Ongoing

Recalled by Insulet Corporation — Acton, MA

Recall Details

Product Type
Devices
Report Date
October 22, 2025
Initiation Date
September 11, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
174,013 units

Product Description

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Reason for Recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Distribution Pattern

US Nationwide distribution.

Code Information

Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X