Z-0132-2026 Class II Ongoing

Recalled by Insulet Corporation — Acton, MA

FDA device recall Z-0132-2026 was initiated by Insulet Corporation on Sep 11, 2025 and is designated Class II. Reason for recall: The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while… The recall status is ongoing. Affected quantity: 174,013 units.

Recall details

Product Type
Devices
Report Date
Oct 22, 2025
Initiation Date
Sep 11, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
174,013 units

Product description

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Reason for recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Distribution pattern

US Nationwide distribution.

Code information

Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0132-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Insulet Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0132-2026. Report a data issue: hello@regdatalab.com.