Z-0591-2024 Class I Ongoing

Recalled by Insulet Corporation — Acton, MA

FDA device recall Z-0591-2024 was initiated by Insulet Corporation on Nov 30, 2023 and is designated Class I. Reason for recall: The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not… The recall status is ongoing. Affected quantity: 27000 users.

Recall details

Product Type
Devices
Report Date
Jan 10, 2024
Initiation Date
Nov 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27000 users

Product description

Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.

Reason for recall

The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Germany and UK.

Code information

UDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0591-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0591-2024. Report a data issue: hello@regdatalab.com.