Z-2016-2025 Class II Ongoing
Recalled by Insulet Corporation
— Acton, MA
FDA device recall Z-2016-2025 was initiated by Insulet Corporation on May 23, 2025 and is designated Class II. Reason for recall: Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual… The recall status is ongoing. Affected quantity: 122,823 downloads.
Recall details
- Product Type
- Devices
- Report Date
- Jul 2, 2025
- Initiation Date
- May 23, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 122,823 downloads
Product description
Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface that controls the Omnipod 5 Automated Insulin Delivery System. It is used to activate/deactivate Pods, display alerts/alarms, and send insulin delivery commands for execution to the Pod. Only available in the US.
Reason for recall
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.
Distribution pattern
US market only- Although there are 122823 downloads of the application, our data shows there are only ~80k active users.
Code information
Model Number: PT-000664/M009-S-AP. UDI-DI: 10385083000336. All software versions 1.X.X
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2016-2025
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Insulet Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works