Z-0154-2023 Class I Terminated

Recalled by Insulet Corporation — Acton, MA

FDA device recall Z-0154-2023 was initiated by Insulet Corporation on Oct 17, 2022 and is designated Class I. Reason for recall: The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint… The recall status is terminated (terminated Sep 17, 2024). Affected quantity: 248,288 US; 167,802 OUS.

Recall details

Product Type
Devices
Report Date
Nov 23, 2022
Initiation Date
Oct 17, 2022
Termination Date
Sep 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
248,288 US; 167,802 OUS

Product description

18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.

Reason for recall

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Distribution pattern

Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

Code information

Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0154-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Insulet Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0154-2023. Report a data issue: hello@regdatalab.com.