Z-0154-2023 Class I Terminated

Recalled by Insulet Corporation — Acton, MA

Recall Details

Product Type
Devices
Report Date
November 23, 2022
Initiation Date
October 17, 2022
Termination Date
September 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
248,288 US; 167,802 OUS

Product Description

18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.

Reason for Recall

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Distribution Pattern

Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

Code Information

Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots