Z-0154-2023 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 23, 2022
- Initiation Date
- October 17, 2022
- Termination Date
- September 17, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 248,288 US; 167,802 OUS
Product Description
18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.
Reason for Recall
The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.
Distribution Pattern
Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.
Code Information
Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots