Z-2164-2020 Class II Terminated

Recalled by Insulet Corporation — Acton, MA

Recall Details

Product Type
Devices
Report Date
June 10, 2020
Initiation Date
April 2, 2020
Termination Date
September 14, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
400

Product Description

Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Reason for Recall

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Distribution Pattern

The products were distributed to the following foreign countries: Italy, Netherlands, UK.

Code Information

UDI: PT - 000010 - 10385082000139; Lot Numbers: L000202