Z-0143-2015 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-0143-2015 was initiated by Merge Healthcare, Inc. on Sep 29, 2014 and is designated Class II. Reason for recall: It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of… The recall status is terminated (terminated Aug 25, 2016). Affected quantity: 52.

Recall details

Product Type
Devices
Report Date
Nov 5, 2014
Initiation Date
Sep 29, 2014
Termination Date
Aug 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52

Product description

Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.

Reason for recall

It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.

Distribution pattern

Nationwide Distribution including the states of AZ, CA, FL, IL, IA, MA, MO, OH, SC, TN, and WI.

Code information

Merge Hemo 9.x with Nellcor SpO2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0143-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0143-2015. Report a data issue: hello@regdatalab.com.