Z-0149-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 5, 2014
- Initiation Date
- September 26, 2014
- Termination Date
- July 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product Description
Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.
Reason for Recall
Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
Distribution Pattern
US Distribution to IL.
Code Information
Part Number: 10641100, SN: 1407006311.