Z-0149-2015 Class II Terminated

Recalled by Tecan US, Inc. — Morrisville, NC

Recall Details

Product Type
Devices
Report Date
November 5, 2014
Initiation Date
September 26, 2014
Termination Date
July 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Reason for Recall

Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.

Distribution Pattern

US Distribution to IL.

Code Information

Part Number: 10641100, SN: 1407006311.