Z-0168-2013 Class II Terminated

Recalled by Alphatec Spine, Inc. — Carlsbad, CA

FDA device recall Z-0168-2013 was initiated by Alphatec Spine, Inc. on Sep 25, 2012 and is designated Class II. Reason for recall: The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws… The recall status is terminated (terminated Mar 29, 2013). Affected quantity: 209 screws.

Recall details

Product Type
Devices
Report Date
Nov 7, 2012
Initiation Date
Sep 25, 2012
Termination Date
Mar 29, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
209 screws

Product description

ILLlCO¿ MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal Pedical Fixation Orthosis, Spondylolisthesis Spinal Fixation The ILLICO MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. When used for a minimally invasive posterior approach ILLlCO¿ MIS Instrumentation is used in conjunction with polyaxial screw components. The ILLICO MIS Posterior Fixation System is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the ILLICO MIS Posterior Fixation System are dependent in part on the configuration of the assembled device and the method of attachment to the spine.

Reason for recall

The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.

Distribution pattern

Nationwide Distribution including AZ, IL, CA, SC, TN, TX, GA, CT, NY, FL, OK, LA, MO, NV,and UT.

Code information

Product Codes: KW, MNI, MNH. Part NumberlLot Number: ¿PART DESCRIPTION LOT 7384S-30 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 30mm 646501 73845-35 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 35mm 646502 73845-40 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 40mm 646503 73845-45 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 45mm 646504 73855-35 Ti CANNULATED POLYAXIAL SCREW 5.5mm X 35mm 646510 73865-30 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 30mm 646515 73865-35 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 35mm 646516 73865-40 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 40mm 646517 73865-45 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 45mm 646542 73865-50 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 50mm 646543 73875-35 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 35mm 646546 73875-45 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 45mm 646548 73875-50 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 50mm 646549

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0168-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Alphatec Spine, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0168-2013. Report a data issue: hello@regdatalab.com.