Z-0169-2013 Class II Terminated

Recalled by Elekta, Inc. — Norcross, GA

FDA device recall Z-0169-2013 was initiated by Elekta, Inc. on Sep 18, 2012 and is designated Class II. Reason for recall: A patient was seriously mistreated after a user made and saved an inadvertent change to the definition of a treatment filed in MOSAIQ. The recall status is terminated (terminated Jun 6, 2013). Affected quantity: 300.

Recall details

Product Type
Devices
Report Date
Nov 7, 2012
Initiation Date
Sep 18, 2012
Termination Date
Jun 6, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
300

Product description

MOSAIQ MOSAIQ is an image-enabled electronic medical record system (EMR) used for oncology workflow management. It lets users supply electronic patient charts and assemble care plans, order diagnostic tests, and prescribe medications. It also lets users import, view, annotate, manipulate, enhance, manage, and archive images.

Reason for recall

A patient was seriously mistreated after a user made and saved an inadvertent change to the definition of a treatment filed in MOSAIQ.

Distribution pattern

Worldwide distribution: USA (nationwide) and country of Canada.

Code information

Version 2.00 to 2.41