Z-0176-2021 Class II Ongoing
FDA device recall Z-0176-2021 was initiated by Bien-Air Surgery Sa on September 11, 2020 and is designated Class II. Reason for recall: The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor. The recall status is ongoing. Affected quantity: Total US = 1120 foot pedals.
Recall Details
- Product Type
- Devices
- Report Date
- October 28, 2020
- Initiation Date
- September 11, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Total US = 1120 foot pedals
Product Description
Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.
Reason for Recall
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Distribution Pattern
US Nationwide distribution including in the states of CA, MI and MN.
Code Information
Model 1600407 / Catalog number 1600407-001 Serial Numbers: 16C0017 16F0013 16I0009 16K0017 17F0039 17I0015 17L0013