Z-0178-2021 Class II Ongoing
FDA device recall Z-0178-2021 was initiated by Bien-Air Surgery Sa on Sep 11, 2020 and is designated Class II. Reason for recall: The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor. The recall status is ongoing. Affected quantity: Total US = 1120 foot pedals.
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2020
- Initiation Date
- Sep 11, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Total US = 1120 foot pedals
Product description
Bien-Air Surgery REF 1600686-001 SN ******* OSSEOSTAP Footcontrol - Product Usage: an active medical device that convert electrical energy into mechanical rotation to a but during Ear, Nose and Throat (ENT) surgery.
Reason for recall
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Distribution pattern
US Nationwide distribution including in the states of CA, MI and MN.
Code information
Model 1600686 / Catalog number 1600686-001 Serial Numbers: 18D0006 18D0007 18D0008 18D0009 18D0010 18D0013 18D0014 18D0015 18D0016 18D0017 18D0018 18D0019 18D0020 18D0021 18D0022 18D0023 18D0024 18D0025 18D0026 18D0027 18D0028 18D0029 18D0030 16A0005 16A0009 16A0010 16A0011 16A0014 17D0011 17G0006
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0178-2021
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for Bien-Air Surgery Sa? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.