Z-0178-2013 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 7, 2012
- Initiation Date
- July 11, 2012
- Termination Date
- June 30, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,602 total for all units
Product Description
Ventlab Premium Small Adult Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.
Reason for Recall
Possible volume leakage through the inlet valve during compressions of the resuscitator bag.
Distribution Pattern
Nationwide Distribution including IL, OH, NY and TN.
Code Information
Product Code: VN500MX, Lot #102194.