Z-0181-2013 Class I Terminated

Recalled by Ventlab Corporation — Mocksville, NC

Recall Details

Product Type
Devices
Report Date
November 7, 2012
Initiation Date
July 11, 2012
Termination Date
June 30, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,602 total for all units

Product Description

AirFlow, Small Adult Resuscitator, AF5000 Series, Single use, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028 The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.

Reason for Recall

Possible volume leakage through the inlet valve during compressions of the resuscitator bag.

Distribution Pattern

Nationwide Distribution including IL, OH, NY and TN.

Code Information

Product Code: AF5140MBPWTD5, Lot #101806 and 102163; AF5140MB-T, Lot #102075 and AFD5140MB-T, Lot #101928 &102211. Recall was amended to include the following: AF5140MBS-I, Lot #101898; AF5140MB-ST, Lot #101896; AF5140MBS, Lot #101900 and AF5140MBS-R, Lot #101899.