Z-0179-2016 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

FDA device recall Z-0179-2016 was initiated by Encore Medical, Lp on Sep 14, 2015 and is designated Class II. Reason for recall: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray… The recall status is terminated (terminated Nov 4, 2016). Affected quantity: 277 units.

Recall details

Product Type
Devices
Report Date
Nov 4, 2015
Initiation Date
Sep 14, 2015
Termination Date
Nov 4, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
277 units

Product description

Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Reason for recall

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Distribution pattern

Nationwide Distribution

Code information

Lot # 52748L02A

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0179-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0179-2016. Report a data issue: hello@regdatalab.com.