Z-0186-2018 Class I Terminated

Recalled by Physio-Control, Inc. — Redmond, WA

FDA device recall Z-0186-2018 was initiated by Physio-Control, Inc. on Oct 27, 2017 and is designated Class I. Reason for recall: The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols. The recall status is terminated (terminated Jun 12, 2019). Affected quantity: 14,219 units in total.

Recall details

Product Type
Devices
Report Date
Dec 20, 2017
Initiation Date
Oct 27, 2017
Termination Date
Jun 12, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,219 units in total

Product description

LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.

Reason for recall

The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.

Distribution pattern

Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.

Code information

713609, 717912, 713904, 718033, 715008, 719323

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0186-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Physio-Control, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0186-2018. Report a data issue: hello@regdatalab.com.