Z-0191-2022 Class II Ongoing

Recalled by FujiFilm Healthcare Americas Corporation — Twinsburg, OH

FDA device recall Z-0191-2022 was initiated by FujiFilm Healthcare Americas Corporation on Oct 14, 2021 and is designated Class II. Reason for recall: Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Nov 10, 2021
Initiation Date
Oct 14, 2021
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3

Reason for recall

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Distribution pattern

Nationwide

Code information

Serial Numbers/UDI: 20514657 (01)04562122958868(11)161220(21)20514657; 20659147 (01)04562122958868(11)20181124(21)20659147; 20659148 (01)04562122958868(11)20181124(21)20659148; 20659149 (01)04562122958868(11)20181124(21)20659149; 20659150 (01)04562122958868(11)20181124(21)20659150; 20659153 (01)04562122958868(11)20181126(21)20659153; 20659154 (01)04562122958868(11)20181126(21)20659154; 20659155 (01)04562122958868(11)20181126(21)20659155; 20659156 (01)04562122958868(11)20181126(21)20659156; 205G0225 (01)04562122958868(11)20170418(21)205G0225; 205G0226 (01)04562122958868(11)20170424(21)205G0226; 205P2146 (01)04562122958868(11)20170615(21)205P2146; 205P2147 (01)04562122958868(11)20170615(21)205P2147; 205P2148 (01)04562122958868(11)20170615(21)205P2148; 205P2149 (01)04562122958868(11)20170615(21)205P2149; 205Q0809 (01)04562122958868(11)20170707(21)205Q0809; 205Q0810 (01)04562122958868(11)20170707(21)205Q0810; 205Q0812 (01)04562122958868(11)20170707(21)205Q0812; 205Q0815 (01)04562122958868(11)20171006(21)205Q0815; 205T7139 (01)04562122958868(11)20171006(21)205T7139; 205T7143 (01)04562122958868(11)20171026(21)205T7143; 205V5544 (01)04562122958868(11)20171026(21)205V5544; 205V5545 (01)04562122958868(11)20171030(21)205V5545; 205V5547 (01)04562122958868(11)20171031(21)205V5547; 205V5548 (01)04562122958868(11)20171031(21)205V5548; G3005239 (01)04562122958868(11)20190308(21)G3005239; G3005240 (01)04562122958868(11)20190308(21)G3005240; G3025041 (01)04562122958868(11)20190603(21)G3025041; G3025042 (01)04562122958868(11)20190604(21)G3025042; G3025044 (01)04562122958868(11)20190605(21)G3025044; G3025045 (01)04562122958868(11)20190605(21)G3025045; G3025048 (01)04562122958868(11)20190606(21)G3025048; G3025049 (01)04562122958868(11)20190606(21)G3025049; G3037183 (01)04562122958868(11)20190624(21)G3037183; G3037184 (01)04562122958868(11)20190624(21)G3037184; G3037192 (01)04562122958868(11)20190709(21)G3037192; G3043580 (01)04562122958868(11)20190820(21)G3043580; G3043581 (01)04562122958868(11)20190820(21)G3043581; G3043582 (01)04562122958868(11)20190820(21)G3043582; G3052217 (01)04562122958868(11)20191016(21)G3052217; G3052218 (01)04562122958868(11)20191016(21)G3052218; G3052225 (01)04562122958868(11)20191202(21)G3052225; G3052234 (01)04562122958868(11)20200128(21)G3052234; G3052236 (01)04562122958868(11)20200109(21)G3052236; G3054831 (01)04562122958868(11)20200115(21)G3054831; G3054833 (01)04562122958868(11)20191212(21)G3054833; G3054840 (01)04562122958868(11)20200115(21)G3054840; G3080616 (01)04562122958868(11)20201215(21)G3080616; G3080617 (01)04562122958868(11)20201215(21)G3080617; G3106256 (01)04562122958868(11)20210413(21)G3106256; G3106258 (01)04562122958868(11)20210413(21)G3106258; G3111572 (01)04562122958868(11)20210413(21)G3111572; G3111573 (01)04562122958868(11)20210421(21)G3111573

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0191-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0191-2022. Report a data issue: hello@regdatalab.com.