Z-0192-2024 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 15, 2023
- Initiation Date
- October 12, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32211 devices
Product Description
EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170
Reason for Recall
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Distribution Pattern
US Nationwide distribution
Code Information
Model BF-1TH190 UDI-DI: 4953170335181; Model BF-H190 UDI-DI: 4953170335174; Model BF-Q170 UDI-DI: 4953170342912 Model BF-Q190 UDI-DI: 4953170335198