Z-0200-2027 Class II Ongoing
FDA device recall Z-0200-2027 was initiated by Greiner Bio-One North America, Inc. on Jul 29, 2026 and is designated Class II. Reason for recall: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++)… The recall status is ongoing. Affected quantity: 860450.
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2026
- Initiation Date
- Jul 29, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 860450
Product description
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Reason for recall
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Distribution pattern
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Code information
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12