Z-0202-2027 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Indianapolis, IN

FDA device recall Z-0202-2027 was initiated by Beckman Coulter, Inc. on Sep 21, 2026 and is designated Class II. Reason for recall: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result… The recall status is ongoing. Affected quantity: 336 units.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Sep 21, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
336 units

Product description

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Reason for recall

Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.

Distribution pattern

US Nationwide distribution.

Code information

Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0202-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0202-2027. Report a data issue: hello@regdatalab.com.