Z-0206-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 8, 2023
- Initiation Date
- September 12, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 sytems
Product Description
RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B SP3 (5.1.3.60023); (3) RayCare 6A (6.0.0.60553)
Reason for Recall
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.
Distribution Pattern
MD, TN
Code Information
UDI-DI: (1)RayCare 5B SP1 (5.1.1.60246) 0735000201053220220316; (2)RayCare 5B SP3 (5.1.3.60023) 0735000201069320221027; (3) RayCare 6A (6.0.0.60553) 0735000201056320220617