Z-0220-2019 Class II Terminated

Recalled by Nakanishi Inc. — Kanuma, N/A

Recall Details

Product Type
Devices
Report Date
October 31, 2018
Initiation Date
June 1, 2018
Termination Date
April 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0

Product Description

Surgical Handpiece, Model: SGS-E2G Order Code: H185, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

Reason for Recall

Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

Distribution Pattern

U.S. Distribution to states of: CA, FL, GA, IL, WA.; and internationally to: Canada.

Code Information

Model Name: SGS-E2G Order Code: H185 *Note: There were no shipments of SGS-E2G between 2011 and April 2018 to the U.S. Market.