Z-0223-2019 Class II Terminated

Recalled by Nakanishi Inc. — Kanuma, N/A

FDA device recall Z-0223-2019 was initiated by Nakanishi Inc. on June 1, 2018 and is designated Class II. Reason for recall: Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manual… The recall status is terminated (terminated April 28, 2021). Affected quantity: 2419.

Recall Details

Product Type
Devices
Report Date
October 31, 2018
Initiation Date
June 1, 2018
Termination Date
April 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2419

Product Description

Micro Surgery Handpiece SGS, Model: SGS-ES, Order Code: H264, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

Reason for Recall

Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

Distribution Pattern

U.S. Distribution to states of: CA, FL, GA, IL, WA.; and internationally to: Canada.

Code Information

Previous Operation Manuals: OM-SH0015E002,OM-SH0015E003, OM-SH0015E004, OM-SH0015E005, OM-SH0100F, OM-SH0096S Model Name: SGS-ES; Order Code H264/H264007 Serial Nos.:0LZ00063 - 0LZ00064, 0LZ00066 - 0LZ00068, 0LZ00070 - 0LZ00109; 0LZ00110 - 0LZ00115; 0BB20112 -0BB20116, 0BB20226 - 0BB20229, 0BB40071 - 0BB40075, 0BB50186 - 0BB50189, 0BB50215 - 0BB50224, 0BB50263, 0BB50265, 0BB50266, 0BB50185, 0BB50190 - 0BB50192; 0BB60051 - 0BB60055, 0BB60079 - 0BB60100, 0BB60106 -0BB60118, 0BB70066 - 0BB70100, 0BB90004 - 0BB90058, 0BB90123 - 0BB90132, 0BBX0168 - 0BBX0175;0BBY0188 -0BBY0192; 0BC10095 - 0BC10099, 0BC10101 - 0BC10110, 0BC10121 - 0BC10130, 0BC10169 - 0BC10170, 0BC10176- 0BC10179, 0BC50093 - 0BC50102, 0BC60001 - 0BC60040, 0BC60085 - 0BC60104, 0BC60078 - 0BC60081, 0BC90001 - 0BC90090; 0BCY0003 - 0BCY0032; 0BD40011 -0BD40020, 0BD60001 - 0BD60034, 0BD60008 - 0BD60010, 0BD60016, 0BD60142-0BD60150, 0BD70062-0BD70072, 0BD80001 -0BD80034, 0BD80095 - 0BD80100, 0BDX0021 -0BDX0040, 0BDX0084- 0BDX0093; 0BDY0001- 0BDY0068, ABE10092 - ABE10121, ABE30001 - ABE30068, ABE30079 - ABE30088, ABE 60001- ABE60068, ABE 60056-ABE60065, ABE60005, ABE60011-ABE60015, ABE60030, ABE60136 - ABE60145, ABE70051- ABE70056, ABE90099-ABE90100; ABEX0001- ABEX0008; ABEY0052 - ABEY0071, ABEX0141 - ABEX0149; ABEY0011 - ABEY0020, ABEY0051; ABF10011- ABF10030, ABF20050, ABF20071 - ABF20079, ABF30014 - ABF30033, ABF70161 - ABF70166; ABFX0046 - ABFX0051, ABFX0053, ABFX0063, ABFY0031 -ABFY0033, ABFY0038- ABFY0040, ABFZ0043-ABFZ0062, ABFY0121 - ABFY0130; ABFZ0001 - ABFZ0040;ABG60147- ABG60154, ABG60156, ABG60159 - ABG60166; ABG70011- ABG70018, ABG70039 - ABG70048, ABG70059 -ABG70090; ABH50001 - ABH50009, ABH50010- ABH50400; ABH40106 - ABH40115; ABH60128 - ABH60137; ABH70001 - ABH70505; ABH80329- ABH80338; ABHX0001 - ABHX0343 - ;