Z-0236-2024 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, INc. — Rochester, NY

FDA device recall Z-0236-2024 was initiated by Ortho-Clinical Diagnostics, INc. on Sep 12, 2023 and is designated Class II. Reason for recall: Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could… The recall status is ongoing. Affected quantity: 342 units.

Recall details

Product Type
Devices
Report Date
Nov 15, 2023
Initiation Date
Sep 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
342 units

Product description

VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)

Reason for recall

Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.

Distribution pattern

US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea

Code information

Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0236-2024
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ortho-Clinical Diagnostics, INc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ortho-Clinical Diagnostics, INc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ortho-Clinical Diagnostics, INc. →

Source: openFDA record for Z-0236-2024. Report a data issue: hello@regdatalab.com.