Z-0257-2020 Class II Terminated
Recalled by GE Healthcare, LLC
— Waukesha, WI
FDA device recall Z-0257-2020 was initiated by GE Healthcare, LLC on Sep 16, 2019 and is designated Class II. Reason for recall: Potential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus… The recall status is terminated (terminated Dec 16, 2020). Affected quantity: 43 (27 US, 16 OUS).
Recall details
- Product Type
- Devices
- Report Date
- Nov 6, 2019
- Initiation Date
- Sep 16, 2019
- Termination Date
- Dec 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 43 (27 US, 16 OUS)
Product description
CardioLab/ComboLab Recording Systems
Reason for recall
Potential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.
Distribution pattern
Nationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.
Code information
Model Number: 2088700-002, No UDI Number, Lot Numbers: RXJ19122006TA, RXJ19122002TA, RXJ19122010TA, RXJ19122001TA, RXJ19122005TA, RXJ19122007TA; Model Number: 2088700-004, UDI Number (Lot Numbers): 01008406821196031119050021RXJ19204004TA (RXJ19204004TA), 01008406821196031119040021RXJ19154006TA (RXJ19154006TA), 01008406821196031119050021RXJ19224011TA (RXJ19224011TA), 01008406821196031119050021RXJ19224017TA (RXJ19224017TA), 01008406821196031119050021RXJ19224014TA (RXJ19224014TA), 01008406821196031119040021RXJ19154009TA (RXJ19154009TA), 01008406821196031119040021RXJ19154011TA (RXJ19154011TA), 01008406821196031119040021RXJ19154010TA (RXJ19154010TA), 01008406821196031119040021RXJ19154007TA (RXJ19154007TA), 01008406821196031119050021RXJ19194009TA (RXJ19194009TA), 01008406821196031119040021RXJ19154003TA (RXJ19154003TA), Not Available (RXJ19224002TA), 01008406821196031119050021RXJ19224016TA (RXJ19224016TA), 01008406821196031119050021RXJ19204003TA (RXJ19204003TA), 01008406821196031119040021RXJ19154001TA (RXJ19154001TA), 01008406821196031119050021RXJ19194003TA (RXJ19194003TA), 01008406821196031119050021RXJ19224013TA (RXJ19224013TA), 01008406821196031119050021RXJ19194002TA (RXJ19194002TA), 01008406821196031119050021RXJ19194011TA (RXJ19194011TA), 01008406821196031119050021RXJ19204007TA (RXJ19204007TA), 01008406821196031119040021RXJ19154005TA (RXJ19154005TA), 01008406821196031119050021RXJ19204006TA (RXJ19204006TA), 01008406821196031119050021RXJ19194007TA (RXJ19194007TA), 01008406821196031119050021RXJ19194004TA (RXJ19194004TA), 01008406821196031119040021RXJ19154008TA (RXJ19154008TA), 01008406821196031119050021RXJ19204002TA (RXJ19204002TA), 01008406821196031119050021RXJ19224001TA (RXJ19224001TA), 01008406821196031119050021RXJ19194001TA (RXJ19194001TA), 01008406821196031119050021RXJ19224012TA (RXJ19224012TA), 01008406821196031119050021RXJ19194008TA (RXJ19194008TA), 01008406821196031119050021RXJ19194005TA (RXJ19194005TA), 01008406821196031119040021RXJ19154004TA (RXJ19154004TA), 01008406821196031119050021RXJ19194006TA (RXJ19194006TA), 01008406821196031119050021RXJ19224009TA (RXJ19224009TA), 01008406821196031119050021RXJ19204005TA (RXJ19204005TA), 01008406821196031119050021RXJ19194010TA (RXJ19194010TA), 01008406821196031119040021RXJ19154002TA (RXJ19154002TA)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0257-2020
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.