Z-0281-2019 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-0281-2019 was initiated by Leica Microsystems, Inc. on Sep 17, 2018 and is designated Class II. Reason for recall: The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound. The recall status is terminated (terminated Aug 21, 2020). Affected quantity: 876 devices (in U.S. and Canada).

Recall details

Product Type
Devices
Report Date
Nov 7, 2018
Initiation Date
Sep 17, 2018
Termination Date
Aug 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
876 devices (in U.S. and Canada)

Product description

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Reason for recall

The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.

Distribution pattern

Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.

Code information

Zebra Printer Model TLP3842 - Affected PSU Part Number 808113-001 or P1028888-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week); Zebra Printer Model GX4301 - Affected PSU Part Number 808101-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0281-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Leica Microsystems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Leica Microsystems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Leica Microsystems, Inc. →

Source: openFDA record for Z-0281-2019. Report a data issue: hello@regdatalab.com.