Z-0283-2018 Class II Terminated

Recalled by Implant Direct Sybron Manufacturing, LLC — Westlake Village, CA

FDA device recall Z-0283-2018 was initiated by Implant Direct Sybron Manufacturing, LLC on Nov 15, 2017 and is designated Class II. Reason for recall: The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm… The recall status is terminated (terminated Oct 9, 2018). Affected quantity: 560 units (worldwide).

Recall details

Product Type
Devices
Report Date
Dec 27, 2017
Initiation Date
Nov 15, 2017
Termination Date
Oct 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
560 units (worldwide)

Product description

Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Reason for recall

The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.

Distribution pattern

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.

Code information

Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0283-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Implant Direct Sybron Manufacturing, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Implant Direct Sybron Manufacturing, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Implant Direct Sybron Manufacturing, LLC →

Source: openFDA record for Z-0283-2018. Report a data issue: hello@regdatalab.com.