Z-0292-2013 Class I Terminated

Recalled by Ethicon, Inc. — Somerville, NJ

FDA device recall Z-0292-2013 was initiated by Ethicon, Inc. on Aug 21, 2012 and is designated Class I. Reason for recall: SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging. The recall status is terminated (terminated Sep 11, 2014). Affected quantity: 28,276.

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Aug 21, 2012
Termination Date
Sep 11, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28,276

Product description

ETHICON SURGIFLO HEMOSTATIC MATRIX KIT Plus FlexTip with Thrombin distributed by ETHICON ETHICON Inc. Somerville, NJ 08876 USA Ref 2993 SURGIFLO Hemostatic Matrix, mixed with sterile saline or thrombin solution, is indicated for surgical procedures (except ophthalmic) for hemostasis, when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical. This is a topical absorbable hemostat consisting of a flowable gelatin matrix (paste) and a lyophilized Thrombin component. This kit also contains components necessary for preparing and applying the matrix during surgery.

Reason for recall

SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Russia, China, and the European Union.

Code information

PMA: P990004 Medical Device Listing Number: D100622 Product Code 2993 Lot/Batch Numbers: 239346 Exp Jan-13 239411 Exp Jun-13

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0292-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0292-2013. Report a data issue: hello@regdatalab.com.