Z-0302-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 4, 2020
- Initiation Date
- October 8, 2020
- Termination Date
- November 17, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 30 units
Product Description
Each KWIK-STIK unit contains a lyophilized microorganism pellet, an ampoule of hydrating fluid, and an inoculating swab. Each device is sealed within a laminated pouch that contains a desiccant to prevent adverse moisture accumulation. KWIK-STIK microorganisms are 3 passages or fewer from the reference culture and are guaranteed to recover when processed using the recommended media and incubation requirements. Available in packs of 2 or 6.
Reason for Recall
KWIK-STIK 2 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145P) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.
Distribution Pattern
Distribution in US, Brazil, Canada, Costa Rica, France, Germany, Ireland, Italy, Japan, New Zealand, Sweden, and United Kingdom.
Code Information
Catalog 01145P, Lots 1145-06-1, 1145-06-2, 1145-06-3, 1145-06-4, 1145-06-6, 1145-07-1 UDI: UDI: 20845357036807