Z-0308-2019 Class II Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

FDA device recall Z-0308-2019 was initiated by Philips Medical Systems (Cleveland) Inc on Jun 7, 2017 and is designated Class II. Reason for recall: It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple… The recall status is terminated (terminated Jul 3, 2019). Affected quantity: 1890.

Recall details

Product Type
Devices
Report Date
Nov 7, 2018
Initiation Date
Jun 7, 2017
Termination Date
Jul 3, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1890

Product description

Ingenuity CT (Model No. 728326), software version 4.1.6 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Reason for recall

It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.

Distribution pattern

US Nationwide Distribution

Code information

software versions 4.1.6

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0308-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems (Cleveland) Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems (Cleveland) Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems (Cleveland) Inc →

Source: openFDA record for Z-0308-2019. Report a data issue: hello@regdatalab.com.