Z-0325-2018 Class II Terminated

Recalled by Amendia, Inc — Marietta, GA

FDA device recall Z-0325-2018 was initiated by Amendia, Inc on July 7, 2017 and is designated Class II. Reason for recall: Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length. The recall status is terminated (terminated October 15, 2019). Affected quantity: 41 screws.

Recall Details

Product Type
Devices
Report Date
January 17, 2018
Initiation Date
July 7, 2017
Termination Date
October 15, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
41 screws

Product Description

Mac Pin Non-Cannulated Screw, 6.5 x 60mm

Reason for Recall

Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length.

Distribution Pattern

Distribution to Georgia, USA.

Code Information

Model #85-6560-1, Lot #132117